13. Set second step’s width time. Input value and press Enter key to
confirm.
Step 002 width
= 5S
14. Set parameters for steps 3 through 5 in the same manner described
above.
15. Set number of run cycles. Input value and press Enter to confirm.
Repeat Count
= 1
16. Select memory position to save list file. Input value and press Enter
to confirm.
CH01
Save list file
= 1
17. Press key to select <Function Mode> and press Enter. Select
<List> and press Enter to confirm.
Fixed
List
18. Press Esc key twice to exit menus.
19. Press On/Off key to turn on the electronic load’s input.
CH01 CC ON
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
20. Press Trig key to trigger List mode operation.